| Low Voltage | High Voltage | |
|---|---|---|
| Operating voltage | 180 - 253 VAC | 342 - 528 VAC |
| Line frequency | 50 / 60 Hz | |
| Phase | 3 | |
| Nominal input power | 5.8 kW | |
| Rated full load / locked rotor current | 19 / 85 | 10 / 42 |
| Minimum electrical service | 30 Amps | 20 Amps |
| Maximum inlet temperature | 90°F (32°C) | |
| Minimum inlet temperature | 50°F (10°C) | |
| Flow rate | 2.75 ± 1.25 gpm (10.4 ± 4.7 lpm) | |
| Maximum inlet pressure | 100 psi (6.9 bars) | |
| Alkalinity | 6.0 - 8.0 pH | |
| Calcium carbonate | < 75 ppm | |
| Deionized water usage | Do not use above 100,000 ohm-cm resistivity | |
| Input power cable (Customer supplied) | 600 VAC 10 Gauge, 3 conductor wire with ground Must conform to local electrical codes | |
| Ambient operating temperature range | 50 - 100ºF (10 - 38ºC) | |
| Gas supply connector Gas return connector EMO remote control receptacle | 1/2 in. Aeroquip self-sealing coupling, male thread. 1/2 in. Aeroquip self-sealing coupling, female thread. Mates with P5 connector part number MS3106A | |
| Adsorber service schedule | 3 Years | |
| Weight | 210 lbs / 95 kg | 222 lbs / 102 kg |
This product has been refurbished by PTB Sales